Direct electrical stimulation using a battery-operated device for induction and modulation of colonic contractions in pigs.

Bertschi, Mattia; Schlageter, Vincent; Vesin, Jean-Marc; Aellen, Steve; Peloponissios, Nicolas; D'Ambrogio, Aris; Wiesel, Paul Herman; Givel, Jean-Claude et al. · Ann Biomed Eng · 2010

basic_science · Level V

Where this comes from

Abstract

Direct electrical stimulation of the colon offers a promising approach for the induction of propulsive colonic contractions by using an implantable device. The objective of this study was to assess the feasibility to induce colonic contractions using a commercially available battery-operated stimulator (maximum pulse width of 1 ms and maximum amplitude of 10 V). Three pairs of pacing electrodes were inserted into the cecal seromuscular layer of anesthetized pigs. During a first set of in vivo experiments conducted on six animals, a pacing protocol leading to cecum contractions was determined: stimulation bursts with 1 ms pulse width, 10 V amplitude (7-15 mA), 120 Hz frequency, and 30-s burst duration, repeated every 2-5 min. In a second testing phase, an evaluation of the pacing protocol was performed in four animals (120 stimulation bursts in total). By using the battery-operated stimulator, contractions of the cecum and movement of contents could be induced in 92% of all stimulations. A cecal shortening of about 30% and an average intraluminal pressure increase of 10.0 +/- 6.0 mmHg were observed.