Transesophageal echocardiography: a simple method for monitoring the patency of ventriculoatrial shunts. Technical note.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 2033439.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A new method for evaluating the patency of a ventriculoatrial shunt is described, and early experience with it is reported. Transesophageal echocardiography can demonstrate a cerebrospinal fluid leak in the right atrium through the atrial tip of a shunting device. This capability was an incidental discovery, and since then the accuracy of the technique in evaluating the patency of a ventriculoatrial shunt has been prospectively studied in 20 observations of 16 patients. The method proved to be accurate in 90% to 100% of cases. It is concluded that transesophageal echocardiography offers a rapid and accurate assessment of ventriculoatrial shunt function, is well tolerated, and is easy to perform.
Medical subject headings
- Cerebrospinal Fluid Shunts
- Echocardiography