Model-based tracking of the hip: implications for novel analyses of hip pathology.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 20347253.
- Also identified by DOI 10.1016/j.arth.2009.12.004.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study investigated the efficacy of a combined high-speed, biplane radiography and model-based tracking technique to study hip joint kinematics and arthrokinematics. Comparing model-based tracking to the gold standard of radiostereometric analysis using implanted metal beads, joint translation was measured with a bias of 0.2 mm and a precision of 0.3 mm, whereas joint rotation was measured with a bias of 0.2° and a precision of 0.8°. A novel measure of hip arthrokinematics characterizing the region of closest contact in the anterosuperior acetabulum was measured with a bias of 0.9% and a precision of 2.5%. Model-based tracking of the hip thus provides the opportunity to noninvasively study hip pathologic conditions such as osteoarthritis and femoroacetabular impingement with great accuracy.
Medical subject headings
- Computer Simulation
- Hip Joint
Anatomy
- hip