Determination of reference geometry for polyethylene tibial insert wear analysis.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 20347254.
- Also identified by DOI 10.1016/j.arth.2010.01.096.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Geometric wear analysis techniques require unworn geometries to serve as a reference in wear measurement. A method to create a reference geometrical model is described for retrieval studies when the actual unworn geometry is unavailable. Never-implanted tibial inserts were scanned with micro-computed tomography. Two, 3, or 6 insert surfaces were coaligned and averaged to create reference geometries. Individual inserts were compared with each other (manufacturing variability) and with the reference geometries (reference variability). The 3-dimensional deviations between the surfaces were recorded. The reference variability was reduced to 8.3 ± 39 μm, vs manufacturing variability of 15 ± 59 μm. Deviations were smallest on the articular surfaces where most wear occurs and were significantly less than the reported insert wear rate of 20 μm/y.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Prosthesis
- Materials Testing
- Models, Theoretical
- Polyethylene
- Prosthesis Design