Free-hand versus novel specialised jig guidance for the passing of intramedullary wires in olecranon fracture fixation: a comparative study.
other
Where this comes from
- Record sourced from PubMed, PMID 20356591.
- Also identified by DOI 10.1016/j.injury.2010.02.034.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The treatment of olecranon fractures frequently involves the use of tension-band fixation. Although associated with high union rates, this method has a high incidence of morbidity associated with soft tissue compromise and limitation of range of movement requiring frequent re-operation for removal of metal. We describe the use of a simple jig to ensure intramedullary placement of longitudinal K-wires and compare the accuracy of placement of Kirschner (K)-wires using this device with the traditional free-hand method. We found the distance from the centre of the medullary canal, the range and standard deviations of K-wire positions to be significantly more precise when the jig was used. This has applications for the use of the device, both with standard metallic radio-opaque wires and potentially with bio-absorbable pins.
Medical subject headings
- Bone Wires
- Fracture Fixation, Internal
- Olecranon Process
- Ulna Fractures