Small photon beam measurements using radiochromic film and Monte Carlo simulations in a water phantom.

García-Garduño, Olivia A; Lárraga-Gutiérrez, José M; Rodríguez-Villafuerte, Mercedes; Martínez-Dávalos, Arnulfo; Celis, Miguel A · Radiother Oncol · 2010

basic_science · Level V

Where this comes from

Abstract

This work reports the use of both GafChromic EBT film immersed in a water phantom and Monte Carlo (MC) simulations for small photon beam stereotactic radiosurgery dosimetry. Circularly collimated photon beams with diameters in the 4-20 mm range of a dedicated 6 MV linear accelerator (Novalis, BrainLAB, Germany) were used to perform off-axis ratios, tissue maximum ratios and total scatter factors measurements, and MC simulations. GafChromic EBT film data show an excellent agreement with MC results (<2.7%) for all measured quantities.