Evaluation of stimulus parameters and electrode geometry for an effective suprachoroidal retinal prosthesis.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 20479521.
- Also identified by DOI 10.1088/1741-2560/7/3/036008.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Several approaches have been proposed for placement of retinal prostheses: epiretinal, subretinal and suprachoroidal. We aimed to systematically evaluate the effectiveness of varying a range of stimulus parameters and electrode geometry for a suprachoroidal electrode array, using cortical evoked responses to monopolar electrical stimulation in cats. Our results indicate that charge thresholds were not dependent on electrode size, pulse widths or position of the return electrode tested, but were dependent on the number of sites stimulated in parallel. Further, we found that the combination of monopolar stimulation with large diameter electrodes, wide pulse widths and parallel stimulation minimized the voltage requirements for stimulation. These results provide useful insights for the design specifications of a low voltage suprachoroidal stimulator.
Medical subject headings
- Choroid
- Electric Stimulation
- Evoked Potentials, Visual
- Retina
- Visual Cortex