Deviations in delineated GTV caused by artefacts in 4DCT.
other
Where this comes from
- Record sourced from PubMed, PMID 20570002.
- Also identified by DOI 10.1016/j.radonc.2010.04.019.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Four-dimensional computed tomography (4DCT) is used for breathing-adapted radiotherapy planning. Irregular breathing, large tumour motion or interpolation of images can cause artefacts in the 4DCT. This study evaluates the impact of artefacts on gross tumour volume (GTV) size. In 19 4DCT scans of patients with peripheral lung tumours, GTV was delineated in all bins. Variations in GTV size between bins in each 4DCT scan were analysed and correlated to tumour motion and variations in breathing signal amplitude and breathing signal period. End-expiration GTV size (GTVexp) was considered as reference for GTV size. Intra-session delineation error was estimated by re-delineation of GTV in eight of the 4DCT scans. In 16 of the 4DCT scans the maximum deviations from GTVexp were larger than could be explained by delineation error. The deviations were largest in the bins adjacent to the end-inspiration bin. The coefficient of variation of GTV size was significantly correlated to tumour motion in the cranio-caudal direction, but no significant correlation was found to breathing signal variations. We found considerable variations in GTV size throughout the 4DCT scans. Awareness of the error introduced by artefacts is important especially if radiotherapy planning is based on a single 4DCT bin.
Medical subject headings
- Artifacts
- Carcinoma, Non-Small-Cell Lung
- Four-Dimensional Computed Tomography
- Lung Neoplasms
- Radiotherapy Planning, Computer-Assisted