Trunnion fracture of the anatomic medullary locking a plus femoral component.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 20570100.
- Also identified by DOI 10.1016/j.arth.2010.03.028.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Fracture of the neck of the femoral component after total hip arthroplasty is an infrequent complication. We present a report of 2 cases of trunnion fractures of fully porous-coated femoral stems (AML A Plus, DePuy International, Leeds, England), which had been implanted for 6 and 7 years after the index procedures. In both cases, failure emanated from a region where the trunnion had been intentionally grooved; and in both cases, the diameter of the trunnion was small, and evidence of corrosion contributing to the initiation of the failure was noted. Although this is the first report of trunnion fractures in this particular implant design, a close monitoring of prostheses with this specific trunnion should be considered. In addition, further analysis of clinical results and complications with this implant is warranted.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Femur
- Hip Prosthesis
- Osteoarthritis, Hip
- Prosthesis Failure
Anatomy
- femur
- hip