Caution required when using an intraoperative knee balancer in total knee arthroplasty.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 20580189.
- Also identified by DOI 10.1016/j.arth.2009.12.014.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Various knee balancers have been developed to assist surgeons in balancing the flexion and extension spaces during total knee arthroplasty. Devices typically are designed to measure the distance between spaces during total knee arthroplasty at a given force level. We analyzed the accuracy of one type of these devices. We found that the devices accurately measure distances, but are quite inaccurate at demonstrating reproducible forces.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Intraoperative Care
Anatomy
- knee