Preferential superior surface motion in wear simulations of the Charité total disc replacement.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 20582556.
- Also identified by DOI 10.1007/s00586-010-1490-y and PMC identifier 3377802.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Laboratory wear simulations of the dual-bearing surface Charité total disc replacement (TDR) are complicated by the non-specificity of the device's center of rotation (CoR). Previous studies have suggested that articulation of the Charité preferentially occurs at the superior-bearing surface, although it is not clear how sensitive this phenomenon is to lubrication conditions or CoR location. In this study, a computational wear model is used to study the articulation kinematics and wear of the Charité TDR. Implant wear was found to be insensitive to the CoR location, although seemingly non-physiologic endplate motion can result. Articulation and wear were biased significantly to the superior-bearing surface, even in the presence of significant perturbations of loading and friction. The computational wear model provides novel insight into the mechanics and wear of the Charité TDR, allowing for better interpretation of in vivo results, and giving useful insight for designing future laboratory physical tests.
Medical subject headings
- Equipment Failure Analysis
- Finite Element Analysis
- Lumbar Vertebrae
- Models, Biological
- Total Disc Replacement
Anatomy
- lumbar spine