Nickel-titanium wire in circumferential suture of a flexor tendon repair: a comparison to polypropylene.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 20610061.
- Also identified by DOI 10.1016/j.jhsa.2010.04.018.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Nickel-titanium (NiTi) has been proposed as an alternative material for flexor tendon core suture. To our knowledge, its suitability as a circumferential suture of flexor tendon repair has not been investigated before. The purpose of this ex vivo study was to investigate the biomechanical properties of NiTi circumferential repairs and to compare them with commonly used polypropylene. Forty porcine flexor tendons were cut and repaired by simple running or interlocking mattress technique using 100 microm NiTi wire or 6-0 polypropylene. The NiTi circumferential repairs showed superior stiffness, gap resistance, and load to failure when compared to polypropylene repairs with both techniques. Nickel-titanium wire seems to be a potential material for circumferential repair of flexor tendons.
Medical subject headings
- Materials Testing
- Nickel
- Polypropylenes
- Sutures
- Tendons
- Titanium