Treatment planning for ferromagnetic seed heating.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2061119.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Computer simulations of ferromagnetic seed heating have been performed for the case of a single constant-temperature ferromagnetic seed. Results obtained from both 2-dimensional and 3-dimensional thermal models are compared to determine the appropriate use and limitations of the 2-dimensional model for the case of array implants. It has been determined that the 2-dimensional model that assumes infinite extent of the ferromagnetic seed(s) is applicable for calculating the thermal distribution in any plane perpendicular to the axes of the implant(s) up to 1 cm from the ends of the seed(s). The feasibility of spatially interspersed ferromagnetic and radioactive seeds for concurrent hyperthermia and brachytherapy has also been studied. The steady-state temperature distribution within an array of ferromagnetic seed implant has been examined as a function of implant spacing and blood perfusion rate. The results have been quantified in generalized histograms that are readily applicable to pre-treatment planning.
Medical subject headings
- Computer Simulation
- Ferric Compounds
- Hyperthermia, Induced