Dosimetry characteristics of metallic cones for intraoperative radiotherapy.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2061127.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Dosimetry data were obtained on the first dedicated linear accelerator of its type designed for electron intraoperative radiotherapy (IORT) within an operating room. The linear accelerator uses a high dose rate, 9 Gy.min-1, to reduce the treatment time. Its chrome-plated brass treatment cones, designed with straight ends and 22.5 degrees beveled ends, are not mechanically attached to the collimator head, but are aligned using a laser projection system. Dosimetry measurements were made for each combination of energy (6, 9, 12, 15, and 16 MeV), cone size (diameters range from 5 to 12 cm), and cone type (22.5 degrees beveled or straight). From these data, depth-dose curves, cone output, and air-gap correction factors were generated that allow the calculation of the monitor setting for delivering a prescribed dose at any depth for any irradiation condition (energy, cone, air gap). Isodose data were measured for every cone using film in a solid water phantom. Scatter off the inside wall of the cone resulted in peripheral dose horns near the surface that were energy and cone dependent, being as large as 120%.
Medical subject headings
- Electrons
- Operating Rooms
- Particle Accelerators