Intramedullary nailing vs. palmar locked plating for unstable dorsally comminuted distal radius fractures: a biomechanical study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 20615595.
- Also identified by DOI 10.1016/j.clinbiomech.2010.06.004.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to compare the stability of a 2.4mm palmar locking compression plate and a new intramedullary nail-plate-hybrid Targon DR for dorsally comminuted distal radius fractures. An extraarticular 10mm dorsally open wedge osteotomy was created in 8 pairs of fresh frozen human radii to simulate an AO-A3-fracture. The fractures were stabilized using one of the fixation methods. The specimens were loaded axially with 200 N and dorsal-excentrically with 80 N. 2000cycles of dynamic loading and axial loading-to-failure were performed. Axial loading revealed that intramedullary osteosynthesis (Targon DR: 369 N/mm) was significantly (p=0.017) stiffer than plate osteosynthesis (Locking compression plate: 131 N/mm). With 214 N/mm the intramedullary nail also showed higher stability during dorsal excentric loading than the Locking compression plate with 51 N/mm (p=0.012). After 2000 cycles of axial loading with 80 N the Targon DR-group was significantly stiffer than the Locking compression plate-group under both loading patterns. Neither group showed significant changes in stiffness after 2000 cycles. Under dorsal excentric loading the Targon DR-group was still significantly stiffer with 212 N/mm than the Locking compression plate-group with 45 N/mm (p=0.012). The load to failure tests demonstrated higher stability of intramedullary nailing (625 N) when compared to plate osteosynthesis (403 N) (p<0.025). The study shows that intramedullary fixation of a distal AO-A3 radial fracture is biomechanically more stable than volar fixed-angle plating under axial and dorsal-excentric loading in an experimental setup.
Medical subject headings
- Bone Plates
- Fracture Fixation, Internal
- Fracture Fixation, Intramedullary
- Fractures, Comminuted
- Radius Fractures
Anatomy
- radius