Depth-resolved blood oxygen saturation assessment using spectroscopic common-path Fourier domain optical coherence tomography.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 20639176.
- Also identified by DOI 10.1109/TBME.2010.2058109 and PMC identifier 3001298.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Although spectroscopic optical coherence tomography (OCT) has been shown to be a promising method for measuring blood oxygen saturation with high-spatial resolution and accuracy, there are several technical issues that need to be addressed before it could become a practical method. In this letter, we have attempted to address two issues that could significantly improve the quantitative assessment of blood oxygen saturation level. First, we have implemented a spectral normalization technique to eliminate the spectral modulation induced by the wavelength-distance-dependent point spread function (PSF) of OCT's. Second, to reduce the spectral speckle noise due to the highly scattering blood, we have implemented a spatial low-pass filter to the 2-D OCT dataset consisting of spectra obtained at different lateral positions. We have assessed the effectiveness of these methods using common-path OCT system. Results showed that we were able to extract unambiguous depth-resolved, SO(2)-dependent spectroscopic information from 1-D and 2-D OCT images, which could be used to accurately assess the SO(2) level.
Medical subject headings
- Fourier Analysis
- Hemoglobins
- Oximetry
- Oxygen
- Oxyhemoglobins
- Tomography, Optical Coherence