Low-dose computed tomography of the paranasal sinus and facial skull using a high-pitch dual-source system--first clinical results.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 20644936.
- Also identified by DOI 10.1007/s00330-010-1892-6.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Computed tomography (CT) of the paranasal sinus is the standard diagnostic tool for a wide range of indications in mostly younger patients. This study aims to assess the image quality of CT of the sinus by using a high-pitch dual-source technique with special regard to the radiation dose. Examinations were performed on a second-generation dual-source CT with a pitch factor of 3.0 (dual-source mode). Images were compared with those with a pitch factor of 0.9 on the same system (single-source mode) and with those of 16-slice CT. Image quality was evaluated by four blinded readers using a 5-point scale (1=poor, 5=excellent). Comparison of the dose length product (DLP) was used to estimate radiation exposure. Seventy-three consecutive patients underwent imaging with the proposed CT protocols. The viewers rated the image quality of the dual-source image sets as nearly as good (3.62) as the single-source images on the same device (4.18) and those on 16-slice CT (3.7). DLP was cut to half of the dose [51 mGycm vs. 97.8 mGycm vs. 116.9 mGycm (p<0.01)]. Using the proposed dual-source mode when examining the paranasal sinus, diagnostic image quality can be achieved while drastically lowering the patient's radiation exposure.
Medical subject headings
- Paranasal Sinuses
- Skull
- Tomography, X-Ray Computed