Factory-assembled polyethylene linings in metal backing show higher wear rates compared to equisized all-polyethylene sockets.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 20647160.
- Also identified by DOI 10.1016/j.arth.2010.05.019.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Two series of cementless total hip arthroplasty with acetabular sockets of a threaded truncated cone design were compared regarding volumetric wear rates. The first series included all-polyethylene acetabular sockets of the type Endler (E-PE); in the second series, a nonmodular titanium metal-backed polyethylene (E-MB) socket with an identical outer shape to E-PE was implanted. Bearings were articulated with alumina 32-mm diameter ball heads. Ninety-five retrieved devices were examined with a modified fluid displacement method using dental self-polymerizing precision casts. The sockets had to be revised mainly because of wear-induced osteolysis: E-PE after 10.6 years on average and E-MB after 7.8 years (P = .002). Comparison with unused sockets showed mean wear rates of 63 mm(3)/y for E-PE and 120 mm(3)/y for E-MB (P = .0008). Increased contact stress and load deformation due to reduction of polyethylene thickness in E-MB compared to E-PE were identified as predominant reasons for higher wear rates.
Medical subject headings
- Acetabulum
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Metals
- Polyethylene
- Prosthesis Design