Management of polyethylene wear associated with a well-fixed modular cementless shell during revision total hip arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 20647163.
- Also identified by DOI 10.1016/j.arth.2010.05.028.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We identified 128 revision total hip arthroplasties from 1993 to 2005 involving a well-fixed Harris-Galante Porous (HGP)-I or HGP-II acetabular component with minimum 2-year post-revision follow-up. Three treatment cohorts were identified. Of the hips that underwent modular liner exchange at revision, 14 hips (25%) required re-revision of the acetabular component, 8 for liner dislodgement, 3 for osteolysis and poly-wear, 2 for dislocation, and 1 for aseptic loosening. Of the hips that underwent revision of the well-fixed shell, 4 (15%) required subsequent re-revision of the acetabular component; 2 for dislocation and 2 for aseptic loosening. Six hips (27%) in the cemented liner group were re-revised; 4 for dislocation and 2 for loosening. Complete revision of a well-fixed HGP acetabular component is more reliable than liner exchange or liner cementation.
Medical subject headings
- Acetabulum
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Polyethylene
- Prosthesis Failure
Anatomy
- hip