A nonradiographic system for assessing pressure for the Codman-Hakim Programmable Valve.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 20679942.
- Also identified by DOI 10.1227/01.NEU.0000380961.26626.BA.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The Codman-Hakim Programmable Valve is widely used in shunting hydrocephalus and other conditions. To establish an accurate valve verification system for the Codman-Hakim programmable valve that does not require radiographic exposure. Prospective clinical trial tested a new valve verification system at 9 research sites. The Valve Programming and Verification System allowed us to establish the valve setting in 62% of subjects with a programmable valve. The Valve Programming and Verification System avoids radiation exposure, is cost-effective, and time efficient for the patient and provider. In approximately one third of subjects, those in whom "adjustment complete" is not achieved, cranial radiographs are still needed.
Medical subject headings
- Cerebrospinal Fluid Shunts
- Hydrocephalus
- Intracranial Pressure
- Software