Integration in a multilayer microfluidic chip of 8 parallel cell sorters with flow control by sol-gel transition of thermoreversible gelation polymer.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 20689871.
- Also identified by DOI 10.1039/c004192k.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Microfluidic systems have significant implications in the field of cell separation since they could provide platforms with inexpensive, disposable and sterile structures. Here, we present a novel strategy to integrate microfluidic sorters into a single chip for high throughput sorting. Our parallel sorter consists of a microfluidic chip with a three-dimensional channel network that utilizes flow switching by a heat-induced sol-gel transition of thermoreversible gelation polymer. The 8 parallel sheathed sample flows were realized by injecting sample and buffer solutions into only 2 inlets. The sheathed flows enabled disposal of unwanted sample waste without laser irradiation, and collection of wanted sample upon irradiation. As an application of the sorter, two kinds of fluorescent microspheres were separated with recovery ratio and purity of 70% or 90% at throughputs of about 100 or 20 particles per second, respectively. Next, Escherichia coli cells expressing green fluorescent protein were separated from those expressing DsRed with recovery ratio and purity of 90% at a throughput of about 20 cells per second.
Medical subject headings
- Cell Culture Techniques
- Cell Separation
- Flow Cytometry
- Flow Injection Analysis
- Microfluidic Analytical Techniques
- Polymers