Reproducibility of 18F-FDG and 3'-deoxy-3'-18F-fluorothymidine PET tumor volume measurements.
Where this comes from
- Record sourced from PubMed, PMID 20720054.
- Also identified by DOI 10.2967/jnumed.110.078501.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The objective of this study was to establish the repeatability and reproducibility limits of several volume-related PET image-derived indices-namely tumor volume (TV), mean standardized uptake value, total glycolytic volume (TGV), and total proliferative volume (TPV)-relative to those of maximum standardized uptake value (SUV(max)), commonly used in clinical practice. Fixed and adaptive thresholding, fuzzy C-means, and fuzzy locally adaptive Bayesian methodology were considered for TV delineation. Double-baseline (18)F-FDG (17 lesions, 14 esophageal cancer patients) and 3'-deoxy-3'-(18)F-fluorothymidine ((18)F-FLT) (12 lesions, 9 breast cancer patients) PET scans, acquired at a mean interval of 4 d and before any treatment, were used for reproducibility evaluation. The repeatability of each method was evaluated for the same datasets and compared with manual delineation. A negligible variability of less than 5% was measured for all segmentation approaches in comparison to manual delineation (5%-35%). SUV(max) reproducibility levels were similar to others previously reported, with a mean percentage difference of 1.8% +/- 16.7% and -0.9% +/- 14.9% for the (18)F-FDG and (18)F-FLT lesions, respectively. The best TV, TGV, and TPV reproducibility limits ranged from -21% to 31% and -30% to 37% for (18)F-FDG and (18)F-FLT images, respectively, whereas the worst reproducibility limits ranged from -90% to 73% and -68% to 52%, respectively. The reproducibility of estimating TV, mean standardized uptake value, and derived TGV and TPV was found to vary among segmentation algorithms. Some differences between (18)F-FDG and (18)F-FLT scans were observed, mainly because of differences in overall image quality. The smaller reproducibility limits for volume-derived image indices were similar to those for SUV(max), suggesting that the use of appropriate delineation tools should allow the determination of tumor functional volumes in PET images in a repeatable and reproducible fashion.
Medical subject headings
- Dideoxynucleosides
- Fluorodeoxyglucose F18
- Positron-Emission Tomography
- Tumor Burden