Apparent motion photometry: evaluation and reliability of a novel method for the measurement of macular pigment.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 20805121.
- Also identified by DOI 10.1136/bjo.2009.178137.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Macular pigment is thought to protect the macula against exposure to light and oxidative stress, both of which may play a role in the development of age-related macular degeneration. The aim was to clinically evaluate a novel cathode-ray-tube-based method for measurement of macular pigment optical density (MPOD) known as apparent motion photometry (AMP). The authors took repeat readings of MPOD centrally (0°) and at 3° eccentricity for 76 healthy subjects (mean (±SD) 26.5±13.2 years, range 18-74 years). The overall mean MPOD for the cohort was 0.50±0.24 at 0°, and 0.28±0.20 at 3° eccentricity; these values were significantly different (t=-8.905, p<0.001). The coefficients of repeatability were 0.60 and 0.48 for the 0 and 3° measurements respectively. The data suggest that when the same operator is taking repeated 0° AMP MPOD readings over time, only changes of more than 0.60 units can be classed as clinically significant. In other words, AMP is not suitable for monitoring changes in MPOD over time, as increases of this magnitude would not be expected, even in response to dietary modification or nutritional supplementation.
Medical subject headings
- Densitometry
- Macula Lutea
- Retinal Pigments