Limitations of imageless computer-assisted navigation for total hip arthroplasty.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 20817389.
- Also identified by DOI 10.1016/j.arth.2010.05.027.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We prospectively evaluated acetabular cup placement in total hip arthroplasty with an imageless computer navigation system or using conventional manual technique. The achieved cup orientation in the manual group had substantially larger variances and greater placement error than the navigation cases. The use of navigation was abandoned in 3 cases because of excessive pelvic tilt and unreliable registration of the pelvis. Computer navigation system helped improve the accuracy of the acetabular cup placement for total hip arthroplasty in this study. The variation between the intraoperative navigation readings and the computed tomographic values suggests that relying on palpation of bony landmarks through drapes and tissue is a limitation of this method. Further, the variation in pelvic tilt has an effect on cup placement that requires further studies.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Imaging, Three-Dimensional
- Osteoarthritis, Hip
- Surgery, Computer-Assisted
Anatomy
- hip
- pelvis