Recurrent dissociation of the tibial insert after mini-subvastus posterior-stabilized total knee arthroplasty: a case report.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 20833051.
- Also identified by DOI 10.1016/j.knee.2010.08.005.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report here on an unusual case of recurrent dissociation of a polyethylene tibial insert from its metal baseplate after high-flex posterior-stabilized (PS) Genesis II total knee arthroplasty with using the mini-subvastus approach. After the 1st episode of dissociation, which had happened 1 month after the primary surgery, we changed the tibial insert and found damages on the dovetails and the post of the retrieved insert. Unfortunately, the patient suffered the same dissociation 1 month after the tibial insert was changed. This case illustrates that incomplete seating of the insert due to limited surgical exposure and anterior impingement of the high-flex PS post may contribute to the risk of dissociation.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Prosthesis
- Prosthesis Design
- Prosthesis Failure
- Tibia