Early type III endoleak with an Endurant endograft.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 20843629.
- Also identified by DOI 10.1016/j.jvs.2010.07.020.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report the case of an 81-year-old man who presented with an intraoperative type III endoleak after treatment with an Endurant endograft for a 60-mm abdominal aortic aneurysm. To our knowledge, this is the first case of a type III endoleak reported with this new device. It was most likely due to a tear in the polyester graft, the cause of which remains speculative. The tear was demonstrated by postoperative angiography, which was more informative than computed tomography. The endoleak was successfully treated by relining with an aorto-uni-iliac device.
Medical subject headings
- Aortic Aneurysm, Abdominal
- Blood Vessel Prosthesis Implantation
- Endoleak
- Endovascular Procedures
- Stents