In vivo oxidation contributes to delamination but not pitting in polyethylene components for total knee arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 20875942.
- Also identified by DOI 10.1016/j.arth.2010.07.010 and PMC identifier 3385511.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to better understand how in vivo oxidation contributes to fatigue damage in total knee arthroplasty (TKA). A total of 119 tibial inserts were consecutively collected after revision surgery. Of the 119 polyethylene retrievals, 29 were gamma sterilized in air (historical), whereas the remaining 90 were gamma sterilized in nitrogen (conventional). Surface damage assessment and characterization of oxidation were performed on all the retrievals. Delamination was significantly more prevalent and extensive in the longer-term, highly oxidized, historical tibial inserts. Pitting damage, in contrast, seemed to be equally prevalent between both retrieval groups and was not correlated with in vivo oxidation. Our findings support our hypothesis that in vivo oxidation is a contributing factor to delamination, but not pitting, in TKA. Despite the lower oxidation displayed by conventional retrievals, this study provides strong evidence that delamination secondary to in vivo oxidation may occur during the second decade of implantation.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Equipment Failure Analysis
- Knee Prosthesis
- Polyethylenes
- Prosthesis Failure
Anatomy
- knee