Thermal damage of the humidified ventilator circuit in the operating room: an analysis of plausible causes.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 20889940.
- Also identified by DOI 10.1213/ANE.0b013e3181ee8092.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Fire in the operating room is a rare but potentially devastating event. In this case report, we describe 2 separate fires of a Westmed heated humidification circuit. We conducted a detailed analysis of potential causes of the fires, including a simulation and series of experiments. Our conclusions were (1) a combination of factors led to the fires; and (2) substantial changes in the design could decrease, but may not completely eliminate, the risk of operating room fire.