Renal artery dissection during repositioning of a mal-deployed aortic endograft with suprarenal fixation.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 20926320.
- Also identified by DOI 10.1016/j.ejvs.2010.08.025.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Endograft mal-deployment during endovascular aneurysm repair (EVAR) may cause renal artery (RA) occlusion. We report bilateral RA occlusion following insertion of an endograft with suprarenal barb fixation. Attempted salvage using the 'pull-down' technique caused right RA dissection. Despite bilateral RA stenting, the right kidney was atrophic on an ultrasound scan at 6 months. This potential hazard of devices with suprarenal barb fixation should be considered when repositioning is attempted.