Audit of an automated checklist for quality control of radiotherapy treatment plans.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 20934764.
- Also identified by DOI 10.1016/j.radonc.2010.09.008.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To assess the effect of adding an automated checklist to the treatment planning process for head and neck intensity-modulated radiotherapy. Plans produced within our treatment planning system were evaluated at the planners' discretion with an automated checklist of more than twenty planning parameters. Plans were rated as accepted or rejected for treatment, during regular review by radiation oncologists and physicists as part of our quality control program. The rates of errors and their types were characterised prior to the implementation of the checklist and with the checklist. Without the checklist, 5.9% of plans were rejected; the use of the checklist reduced the rejection rate to 3.1%. The checklist was used for 64.7% of plans. Pareto analysis of the causes of rejection showed that the checklist reduced the number of causes of rejections from twelve to seven. The use of an automated checklist has reduced the need for reworking of treatment plans. With the use of the checklist, most rejections were due to errors in prescription or inadequate dose distributions. Use of the checklist by planners must be increased to maximise improvements in planning efficiency.
Medical subject headings
- Checklist
- Head and Neck Neoplasms
- Quality Assurance, Health Care
- Radiotherapy Planning, Computer-Assisted