Time-varying autoregressive model-based multiple modes particle filtering algorithm for respiratory rate extraction from pulse oximeter.
other
Where this comes from
- Record sourced from PubMed, PMID 20937577.
- Also identified by DOI 10.1109/TBME.2010.2085437.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We present a particle filtering algorithm, which combines both time-invariant (TIV) and time-varying autoregressive (TVAR) models for accurate extraction of breathing frequencies (BFs) that vary either slowly or suddenly. The algorithm sustains its robustness for up to 90 breaths/min (b/m) as well. The proposed algorithm automatically detects stationary and nonstationary breathing dynamics in order to use the appropriate TIV or TVAR algorithm and then uses a particle filter to extract accurate respiratory rates from as low as 6 b/m to as high as 90 b/m. The results were verified on 18 healthy human subjects (16 for metronome and 2 for spontaneous measurements), and the algorithm remained accurate even when the respiratory rate suddenly changed by 24 b/m (either increased or decreased by this amount). Furthermore, simulation examples show that the proposed algorithm remains accurate for SNR ratios as low as -20 dB. We are not aware of any other algorithms that are able to provide accurate TV BF over a wide range of respiratory rates directly from pulse oximeters.
Medical subject headings
- Algorithms
- Monitoring, Physiologic
- Oximetry
- Respiratory Rate
- Signal Processing, Computer-Assisted