Pre-stenting with a bare metal stent before percutaneous pulmonary valve implantation: acute and 1-year outcomes.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 20965979.
- Also identified by DOI 10.1136/hrt.2010.198382.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine the feasibility and safety of pre-stenting with a bare metal stent (BMS) before percutaneous pulmonary valve implantation (PPVI), and to analyse whether this approach improves haemodynamic outcomes and impacts on the incidence of PPVI stent fractures. Retrospective analysis of prospectively collected data. Tertiary paediatric and adult congenital heart cardiac centre. 108 consecutive patients with congenital heart disease underwent PPVI between September 2005 and June 2008 (54 with PPVI alone, 54 with BMS pre-stenting before PPVI). There were no significant differences in procedural complication rates. Acutely, there was no difference in haemodynamic outcomes. Serial echocardiography revealed that in the subgroups of 'moderate' (26-40 mm Hg) and 'severe' (>40 mm Hg) right ventricular outflow tract (RVOT) obstruction, patients with pre-stenting showed a tendency towards lower peak RVOT velocities compared to patients after PPVI alone (p=0.01 and p=0.045, respectively). The incidence of PPVI stent fractures was not statistically different between treatment groups at 1 year (PPVI 31% vs BMS+PPVI 18%; p=0.16). However, pre-stenting with BMS was associated with a lower risk of developing PPVI stent fractures (HR 0.35, 95% CI 0.14 to 0.87, p=0.024). The probability of freedom from serious adverse follow-up events (death, device explantation, repeat PPVI) was not statistically different at 1 year (PPVI 92% vs BMS+PPVI 94%; p=0.44). Pre-stenting with BMS before PPVI is a feasible and safe modification of the established implantation protocol. Pre-stenting is associated with a reduced risk of developing PPVI stent fractures.
Medical subject headings
- Angioplasty, Balloon, Coronary
- Heart Valve Prosthesis Implantation
- Pulmonary Valve Insufficiency
- Stents