Accidental intraperitoneal tunneling of driveline of left ventricular assist device.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 20971293.
- Also identified by DOI 10.1016/j.athoracsur.2010.04.102.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report a case of accidental intraperitoneal tunneling of the driveline of a left ventricular assist device, which was detected at time of pump exchange. The driveline was completely wrapped with the greater omentum. This technical mistake made during the original left ventricular assist device implantation enabled the patient to remain free from any driveline or pump infection for 4 years and 1 month.
Medical subject headings
- Heart-Assist Devices