High-power potassium-titanyl-phosphate laser fibres for endovaporization of benign prostatic hyperplasia: how much do they deteriorate during the procedure?
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 21040366.
- Also identified by DOI 10.1111/j.1464-410X.2010.09760.x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
• To assess the residual power delivered at the tip of a high-power (80 W) potassium-titanyl-phosphate (KTP) laser (80 W Green Light PV(TM), Laserscope(®); American Medical Systems, Minnetonka, MN, USA) at the end of a photoselective vaporization of the prostate (PVP) procedure, as well as the deflection angle of the laser beam. • In total, 65 laser fibres were collected at the end of PVP procedures indicated for symptomatic benign prostatic hyperplasia over a period of 17 months. PVP was performed by two senior urologists. • The power of laser beam at tip exit was measured for each fibre at the end of the procedures using a photodiode whose signal was amplified, and then quantified by a volt multimeter. • The deflection angle was measured using a graduated sphere. • Approximately 70% of fibres delivered less than 40 W at the end of the procedure, which is less than the vaporization threshold. • Some 9% of fibres had lost their diffraction capacity with a significant alteration of laser beam angulation. These values were not operator dependent. • The data show that a large proportion of laser fibers deliver a significantly underpowered beam at the end of the procedures. • This seems to be caused by peroperative destruction of the fibers, which results in a progressive loss of efficacy of PVP during procedures.
Medical subject headings
- Equipment Failure Analysis
- Laser Therapy
- Lasers, Solid-State
- Prostatectomy
- Prostatic Hyperplasia