Reproducibility of non-standardised autonomic function testing in the pre-operative assessment screening clinic*.
Where this comes from
- Record sourced from PubMed, PMID 21091663.
- Also identified by DOI 10.1111/j.1365-2044.2010.06566.x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
By convention, autonomic function tests are undertaken under standard test conditions that limit their implementation during routine pre-operative assessment. We therefore evaluated the comparability of autonomic function tests under both non-standardised and standardised test conditions in 20 healthy male subjects. Autonomic function was assessed using an ECG monitor and a continuous non-invasive blood pressure measurement device. Under non-standardised conditions, intraclass correlation for heart rate variability analysis was good for the low and high frequency bands (0.87; 95% CI 0.58-0.96 and 0.83; 95% CI 0.56-0.94, respectively), but moderate (0.65; 95% CI 0.14-0.86) for the very low frequency band; reproducibility was high for the expiration/inspiration ratio (0.89; 95% CI 0.71-0.96), Valsalva ratio (0.76; 95% CI 0.37-0.91) and handgrip test (0.76; 95% CI 0.35-0.91) (all p<0.05) but was low for the response to quick standing. Reproducibility under standardised conditions was comparable to the above values. We demonstrated that reproducibility for most autonomic tests under non-standardised conditions is acceptable and suggest that implementation of these tests during pre-operative assessment may be feasible.
Medical subject headings
- Autonomic Nervous System
- Preoperative Care