Predictable PaCO2 with two different flow settings using the Mapleson D system.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 2111629.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Two different settings of fresh gas flow (VFG) and minute ventilation (VE) used with the coaxial Mapleson D system (Bain), were evaluated in 59 adults (ASA I-III) during controlled ventilation and different types of surgical procedures. The two flow settings (alternatives A and B) were VFG of 75 and 110 ml.min-1.kg-1 and VE of 150 and 175 ml.min-1.kg-1, aiming to generate normocapnea and mild hypocapnea, respectively. The PaCO2 obtained with alternative A was 5.5 +/- 0.5 kPa (mean +/- s.d.), with 92% of the patients within the range 4.7-6.1 kPa. With alternative B, the PaCO2 was 4.4 +/- 0.5 kPa, with 82% of the patients within the range 3.5-4.9 kPa. It is concluded that these two flow regimes are suitable for clinical use when either normocapnea or mild hypocapnea is desired.
Medical subject headings
- Anesthesia, Inhalation
- Carbon Dioxide
- Ventilators, Mechanical