MRI-compatible intensity-modulated force sensor for cardiac catheterization procedures.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 21118758.
- Also identified by DOI 10.1109/TBME.2010.2095853.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper presents a novel, magnetic resonance imaging (MRI)-compatible, force sensor suitable for cardiac catheterization procedures. The miniature, fiber-optic sensor is integrated with the tip of a catheter to allow the detection of interaction forces with the cardiac walls. The optical fiber light intensity is modulated when a force acting at the catheter tip deforms an elastic element, which, in turn, varies the distance between a reflector and the optical fiber. The tip sensor has an external diameter of 9 Fr (3 mm) and can be used during cardiac catheterization procedures. The sensor is able to measure forces in the range of 0-0.85 N, with relatively small hysteresis. A nonlinear method for calibration is used and real-time MRI in vivo experiments are carried out, to prove the feasibility of this low-cost sensor, enabling the detection of catheter-tip contact forces under dynamic conditions.
Medical subject headings
- Cardiac Catheterization
- Magnetic Resonance Imaging
- Surgery, Computer-Assisted