Molecular orientation affects localization accuracy in superresolution far-field fluorescence microscopy.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 21133355.
- Also identified by DOI 10.1021/nl103472b.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We investigate the cooperative effect of molecular tilt and defocus on fluorophore localization by centroid calculation in far-field superresolution microscopy based on stochastic single molecule switching. If tilt angle and defocus are unknown, the localization contains systematic errors up to about ±125 nm. When imaging rotation-impaired fluorophores of unknown random orientation, the average localization accuracy in three-dimensional samples is typically limited to about ±32 nm, restricting the attainable resolution accordingly.