Mid-term results of an uncemented femoral stem with modular neck options.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 21157745.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We prospectively assessed the results of 239 primary total hip replacements performed using a conical stem combined with modular necks of different lengths and inclinations (Modulus System, Lima Corporate San Daniele Del Friuli, Udine, Italia) in 222 patients (50 men, 172 women), undergoing surgery between October 2001 and December 2006 and presenting with anatomical deformities of the proximal femur and/or acetabulum, including developmental dysplasia (DDH), ankylosis, and sequelae of osteotomies or fractures. Such conditions can make hip replacement problematic. The mean age at the time of surgery was 57.6 years (22 ÷ 83). No patients were lost to follow-up. 3 femoral components underwent revision. At a mean of 5 years follow-up the Harris Hip Score showed a significant improvement, increasing from 35 preoperatively to a mean of 96.6. Using Kaplan-Meier analysis the survival rate at 5 years was 98.28%. The Modulus stem showed good mid-term results in terms of survival, as well as clinical and radiographic outcome.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Cementation
- Hip Prosthesis
- Prosthesis Design
- Prosthesis Failure
Anatomy
- femur
- hip