A reproducible and inexpensive method of measuring hip abductor strength.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 21157757.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The evaluation of hip abductor strength is useful in assessing of the outcome of hip surgery. Hand-held dynamometers are available, but they are less reliable in assessing hip abductor strength than some other muscle groups. We describe a new device designed to measure hip abductor strength, which is practical in a clinical setting. A system of constraints, pads and reference points was devised to make force measurements as little examiner-dependent as possible. Reproducibility was assessed in a controlled setting. The abductor strength of ten healthy young subjects (average age 28 years) was tested twice on each side by two independent examiners. Tests were performed in a supine position, eliminating the influence of gravity and examiner intervention. The results indicated high reproducibility, the maximal measurement uncertainty being within 1 N. Intra-class correlation coefficients ranged from 0.85-0.98 for intra-rater reproducibility, and 0.81-0.96 for inter-rater reporoducibility. The coefficient of variation was lower than 10%. The device described may be suitable for routine clinical assessment of patients after hip surgery.
Medical subject headings
- Hip
- Muscle Strength
- Muscle Strength Dynamometer
- Muscle, Skeletal
Anatomy
- hip