Implantable ventricular assist device exchange with focused intravascular deairing techniques.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 21172546.
- Also identified by DOI 10.1016/j.athoracsur.2010.04.012.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
As ventricular assist devices are increasingly adopted and widely implemented as a highly effective therapy for end-stage heart disease, extended utilization periods for destination therapy or bridge-to-transplantation have created the possibility of device failure, infection, or thrombosis, requiring challenging implant exchanges. A major problem in these operations is the risk of air embolization, particularly in a nonsternotomy approach that precludes access to the outflow aortic graft and to the ascending aorta. We report a minimally invasive, nonsternotomy HeartMate II implantable left ventricular assist device (LVAD) exchange, using peripheral cardiopulmonary support and a novel approach to continuous intravascular ascending aortic air removal.