Rapid, sensitive analysis of protein mixtures by mass spectrometry.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2118659.
- Also identified by PMC identifier 54640.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We have developed a method for determining the molecular masses of proteins in complex mixtures by mass spectrometry. The method has the capacity to examine the components of mixtures without using any chromatographic separation steps and will tolerate relatively large amounts of buffers and inorganic contaminants. It allows the simultaneous determination of protein molecular masses from 1 to 40 kDa with an accuracy of +/- 0.01% and above 40 kDa with reduced accuracy. The lower limit for practical detection of a protein is a concentration of approximately 0.1 microM, and less than 1 microliter of such a solution is consumed. The analysis is very fast: less than 15 min is necessary to perform the complete analysis, including sample preparation, introduction into the mass spectrometer, mass spectrum collection, and data reduction. The mass spectrum that is obtained does not require elaborate interpretation because there is no fragmentation of the ionized protein (or protein subunit) molecule. Therefore, there is a one-to-one correspondence between the peaks in the mass spectrum and the proteins present in the original mixture. The spectra assume the appearance of chromatograms, with the abscissa being mass-to-charge ratio rather than chromatographic retention time.
Medical subject headings
- Proteins