Near-infrared spectroscopy: a novel, noninvasive, diagnostic method for detrusor overactivity in patients with overactive bladder symptoms--a preliminary and experimental study.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 21216087.
- Also identified by DOI 10.1016/j.eururo.2010.12.032.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Near-infrared spectroscopy (NIRS) is an optical technology. It detects the hemodynamic changes in tissues via noninvasive measurement of changes in the concentration of tissue chromophores such as oxyhemoglobin (O(2)Hb) and deoxyhemoglobin (HHb). Involuntary bladder contractions may cause changes detectable by NIRS. To address the accuracy and reproducibility of NIRS to detect the hemodynamic effects of detrusor overactivity (DO). A prospective cohort study was carried out on 41 patients with overactive bladder symptoms. Forty-one patients underwent one or more filling cystometries with simultaneous NIRS of the bladder. The separated graphs representing both tests were presented to three urodynamicists on two occasions, 3 wk apart. The graphs showed curves with and without DO episodes with the bladder sensations marked. Thirteen of 47 graphs (28%) with DO and 16 of 58 graphs (28%) without DO were excluded due to motion artifacts. The urodynamicists marked pressure changes suggestive of DO on the cystometry curves. For NIRS curves they marked definite deviations from baseline. The sensitivity and specificity of NIRS for DO were determined. The inter- and intraobserver agreements were determined. Valid data from 33 of 41 patients (80%) were included in the analysis. The interobserver agreement to trace the effect of DO on NIRS curves was "substantial" (κ(f)>0.6). The sensitivity of the Hb(sum) (O(2)Hb+HHb) curves for DO was 62-97% with a specificity of 62-79% (area under the curve [AUC]: 0.80-0.82; p<0.001). O(2)Hb curves had 79-85% sensitivity and 82-91% specificity for DO (AUC: 0.80-0.85; p<0.001). The sensitivity and specificity of the HHb curves for DO were 71-82% and 77-82%, respectively (AUC: 0.73-0.84; p<0.001). These values represent the performance of NIRS in the data sample that is not contaminated with motion artifacts; they are not representative of a general clinical setting. NIRS is a potential noninvasive, reproducible, diagnostic method to detect DO.
Medical subject headings
- Hemodynamics
- Spectroscopy, Near-Infrared
- Urinary Bladder
- Urinary Bladder, Overactive