Transcranial electric motor evoked potential monitoring during spine surgery: is it safe?

Schwartz, Daniel M; Sestokas, Anthony K; Dormans, John P; Vaccaro, Alexander R; Hilibrand, Alan S; Flynn, John M; Li, P Mark; Shah, Suken A et al. · Spine (Phila Pa 1976) · 2011

retrospective_cohort · Level III

Where this comes from

Abstract

Retrospective review. To report on the safety of repetitive transcranial electric stimulation (RTES) for eliciting motor-evoked potentials during spine surgery. Theoretical concerns over the safety of RTES have hindered broader acceptance of transcranial electric motor-evoked potentials (tceMEP), despite successful implementation of spinal cord monitoring with tceMEPs in many large spine centers, as well as their apparent superiority over mixed-nerve somatosensory-evoked potentials (SSEP) for detection of spinal cord injury. The records of 18,862 consecutive patients who met inclusion criteria and underwent spine surgery with tceMEP monitoring were reviewed for RTES-related complications. This large retrospective review identified only 26 (0.14%) cases with RTES-related complications; all but one of these were tongue lacerations, most of which were self-limiting. The results demonstrate that RTES is a highly safe modality for monitoring spinal cord motor tract function intraoperatively.

Medical subject headings