A model of waveform reliability in facial nerve electroneurography.

Gutnick, H N; Kelleher, M J; Prass, R L · Otolaryngol Head Neck Surg · 1990

Where this comes from

Abstract

This study identifies four factors that should be considered when evaluating the reliability of new testing methods in facial nerve electroneurography (ENOG): (1) waveform peak-to-peak amplitude; (2) response variability; (3) statistical significance of day-to-day test/retest differences; and (4) waveform morphology. These factors were applied to data gathered on seven normal subjects who had three trials of facial-nerve ENOG on three separate test days in order to determine an optimal recording electrode montage. Both bipolar and monopolar recording-electrode montages were used on the nasal alae and on the nasolabial crease. Resultant peak-to-peak amplitudes of combined motor action potentials were recorded and measured using a clinical evoked-potential instrument. A biphasic waveform shape with optimal peak-to-peak point rather than the nasolabial crease. The contribution to the assessment of response reliability of each of the four factors is discussed.