Transesophageal Doppler ultrasonography: evidence for improved cardiac output monitoring.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 2122771.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Cardiac output monitoring by transesophageal Doppler ultrasound has not gained wide clinical acceptance. A recently developed transesophageal Doppler device, Accucom 2, features technological advances aimed to reduce the error of this approach to the monitoring of cardiac output. To determine if Accucom 2 enables more accurate cardiac output monitoring, a prospective investigation was undertaken. Analyses of 923 simultaneous measurements of transesophageal Doppler and thermodilution cardiac output in 47 anesthetized patients were compared. Results using Accucom 2 were highly correlated with the results obtained with thermodilution (R = 0.91) and provided significantly greater accuracy in comparison with a previously developed Doppler cardiac output monitor, Accucom 1. Technological advances in transesophageal Doppler ultrasound hold promise for clinically useful noninvasive monitoring of cardiac output.
Medical subject headings
- Cardiac Output
- Esophagus
- Monitoring, Physiologic