Quantifying and reducing the effect of calibration error on variability of PET/CT standardized uptake value measurements.
Where this comes from
- Record sourced from PubMed, PMID 21233174.
- Also identified by DOI 10.2967/jnumed.110.083865 and PMC identifier 4491194.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to measure the errors introduced by regular calibration of PET/CT scanners and to minimize the effect of calibration error on standardized uptake value measurements. Global calibration factors from 2 PET/CT scanners were recorded for 3.5 and 1.8 y, comparing manufacturer-recommended protocols with modified protocols to evaluate error contributions due to operator-influenced procedures. Dose calibrator measurements were evaluated using National Institute of Standards and Technology-traceable sources. Dose calibrator variability was less than 1%, although there was a consistent bias. Global scaling variability was reduced from 6% to 4% for scanner 1 and from 11% to 4% for scanner 2 when quality assurance and quality control procedures were applied to the calibration protocol. When calibrations were done using a (68)Ge/(68)Ga phantom, the variability for both scanners was reduced to approximately 3%. Applying quality assurance and quality control procedures to scanner calibration reduces variability, but there is a still a residual longitudinal scanner variability of 3%-4%. The procedures proposed here reduce the impact of operator error on scanner calibration and thereby minimize longitudinal variability in standardized uptake value measurements.
Medical subject headings
- Positron-Emission Tomography
- Tomography, Emission-Computed