Early aseptic loosening with a precoated low-profile tibial component: a case series.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 21236628.
- Also identified by DOI 10.1016/j.arth.2010.11.002.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Between March 2007 and December 2008, we performed 529 consecutive total knee arthroplasties in 460 patients with the Zimmer (Warsaw, IN) NexGen MIS Tibial Component using a minimally invasive approach. Eight knees in 8 patients (1.5%) were revised for early aseptic loosening of the tibial component despite normal initial postoperative radiographs. Several additional patients have concerning radiographic signs of pending failure. The mean time to revision was 17 months (range, 9-31 months). Intraoperatively, in all cases, more than 50% of the tibial tray was devoid of cement and factory-applied polymethylmethacrylate. Our experience with early aseptic loosening of this tibial component has led us to discontinue its use until the etiology of the high early failure rate is able to be determined.