Single versus sequential testing with scanning peripheral anterior chamber depth analyser, IOLMaster and anterior segment optical coherence tomography for the detection of narrow angles.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 21242582.
- Also identified by DOI 10.1136/bjo.2009.175869.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Scanning peripheral anterior chamber depth analyser (SPAC), IOLMaster and anterior segment optical coherence tomography (AS-OCT) are devices that can screen non-invasively for narrow angles (NAs) and may reduce the need for gonioscopy, especially where the availability of ophthalmologists is limited. When used individually to detect NAs, neither SPAC, AS-OCT nor IOLMaster were able to achieve both high sensitivity and specificity when using gonioscopy as a gold standard. To assess whether sequential testing improves discrimination of NAs over single testing. In a prospective community-based cross-sectional study in Singapore, 2047 phakic right eyes underwent SPAC, IOLMaster and AS-OCT. NA eyes were defined as present when Shaffer scores were 0 or 1 in ≥ 2 quadrants on gonioscopy. Sensitivities and specificities of single and sequential testing were compared. At a fixed specificity of 95%, sensitivities in detecting NA eyes were 57.8% (95% CI 52.7% to 62.7%) for SPAC, 43.3% (95% CI 38.3% to 48.3%) for IOLMaster and 52.1% (95% CI 46.4% to 57.7%) for AS-OCT. After sequential testing using SPAC followed by AS-OCT, the sensitivity rose to 70.3% (95% CI 65.5% to 74.8%) while specificity remained at 94.3% (95% CI 93.1% to 95.4%), with sensitivity significantly higher than single testing alone (p<0.001). Sequential testing significantly improves the accuracy of detecting NAs and may be useful in screening settings.
Medical subject headings
- Anterior Eye Segment
- Diagnostic Techniques, Ophthalmological
- Glaucoma, Angle-Closure