Anterior-posterior glide mobile-bearing total knee arthroplasty: complications related to prosthesis design.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 21256701.
- Also identified by DOI 10.1016/j.arth.2010.10.007.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to evaluate complications related to prosthesis design in patients managed with cemented total knee arthroplasty using anterior-posterior glide mobile-bearing prosthesis. One hundred eighty-three total knee arthroplasties were performed using this prosthesis on 146 patients, with a minimum of 7 years of follow-up. There were late dislocations of the bearing in 2 knees, anterior soft tissue impingements in 6 knees at an earlier follow-up, excessive anteroposterior translation in 5 knees, and a fracture of the femoral component induced by osteolysis in 1 knee. Design modifications in the restraint mechanisms of the anterior-posterior glide mobile-bearing prosthesis may be required to minimize the development of complications with regard to sagittal instability and to improve the prosthesis' longevity.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Joint Instability
- Knee Dislocation
- Knee Prosthesis
- Prosthesis Design
Anatomy
- knee